Reference Manual
00809-0100-4530, Rev BA
July 2009
I-3
Rosemount 5300 Series
MAINT_ALM 78 Alarm indicating the device needs maintenance soon. If the condition is ignored, the
device will eventually fail.
MAINT_ENABLE 75 Enabled MAINT_ALM alarm conditions. Corresponds bit for bit to the MAINT_ACTIVE. A
bit on means that the corresponding alarm condition is enabled and will be detected. A bit
off means the corresponding alarm condition is disabled and will not be detected.
MAINT_MASK 76 Mask of MAINT_ALM. Corresponds bit of bit to MAINT_ACTIVE. A bit on means that the
condition is masked out from alarming.
MAINT_PRI 74 Designates the alarming priority of the MAINT_ALM
MANUFAC_ID 10 Manufacturer identification number – used by an interface device to locate the DD file for
the resource.
MAX_NOTIFY 31 Maximum number of unconfirmed notify messages possible.
MEMORY_SIZE 22 Available configuration memory in the empty resource. To be checked before
attempting a download.
MESSAGE_DATE 57 Date associated with the MESSAGE_TEXT parameter.
MESSAGE_TEXT 58 Used to indicate changes made by the user to the device's installation, configuration,
or calibration.
MIN_CYCLE_T 21 Time duration of the shortest cycle interval of which the resource is capable.
MISC_OPTION 47 Indicates which miscellaneous licensing options are enabled.
MODE_BLK 05 The actual, target, permitted, and normal modes of the block:
Target: The mode to “go to”
Actual: The mode the “block is currently in”
Permitted: Allowed modes that target may take on
Normal: Most common mode for actual
NV_CYCLE_T 23 Minimum time interval specified by the manufacturer for writing copies of NV parameters
to non-volatile memory. Zero means it will never be automatically copied. At the end of
NV_CYCLE_T, only those parameters which have changed need to be updated in
NVRAM.
OUTPUT_BOARD_SN 53 Output board serial number.
PWA_SIMULATE 85 Parameter allowing simulation of PWA alarms.
RB_SFTWR_REV_ALL 51 The string will contains the following fields:
Major rev: 1-3 characters, decimal number 0-255
Minor rev: 1-3 characters, decimal number 0-255
Build rev: 1-5 characters, decimal number 0-255
Time of build: 8 characters, xx:xx:xx, military time
Day of week of build: 3 characters, Sun, Mon,...
Month of build: 3 characters, Jan, Feb.
Day of month of build: 1-2 characters, decimal number 1-31
Year of build: 4 characters, decimal
Builder: 7 characters, login name of builder
RB_SFTWR_REV_BUILD 50 Build of software that the resource block was created with.
RB_SFTWR_REV_MAJOR 48 Major revision of software that the resource block was created with.
RB_SFTWR_REV_MINOR 49 Minor revision of software that the resource block was created with.
RECOMMENDED_ACTION 68 Enumerated list of recommended actions displayed with a device alert.
RESTART 16 Allows a manual restart to be initiated. Several degrees of restart are possible. They are
the following:
1 Run – nominal state when not restarting
2 Restart resource – not used
3 Restart with defaults – set parameters to default values. See
START_WITH_DEFAULTS below for which parameters are set.
4 Restart processor – does a warm start of CPU.
RS_STATE 07 State of the function block application state machine.
SAVE_CONFIG_BLOCKS 62 Number of EEPROM blocks that have been modified since last burn. This value will
count down to zero when the configuration is saved.
Parameter
Index
Number
Description
Kommentare zu diesen Handbüchern